MAURER - DIGITAL - 2-COLOR - PYROMETER

Series QKTRD 1075

Temperature range 300 to 1200°C (572 - 2192°F)

Temperature control during production process

compact unit

Infrared - measuring transducer and electronic process unit in one case


QKTRD 1000
  • Vario-objective (vario-focus objective)
  • light beam aiming device with green LED for target
  • emissivity adjustable at the unit
  • 2 - and 1 - color temperature
  • analog- und digital output (serial interface)
  • 1 limit output (open collector)
  • software IR-LOG
Through the serial interface additional parameter functions are possible:
analog output: 0 - 20 / 4 - 20 mA switchable
zoom range within measuring range
emissivity: 100,00 - 10,00 %
average: arithmetical or sliding
maximum value storage: storage modes and erase functions par ex. automatically with the next measuring object
 
 
 
Examples for applications: steel, iron, non-ferrous metal, tempering, hardening, induction heating, laser, forging, vacuum furnace, pre-heating, rolling



Technical datas:  
Unit types QKTRD 1075-1 QKTRD 1075-2
Target marking light beam aiming device green LED or laser viewfinder
Temp. measuring ranges: MR 1: 300 - 900°C (572 - 1652°F)
MR 2: 350 - 1000°C (662 - 1832°F)
MR 3: 400 - 1100°C (752 - 2012°F)
MR 4: 500 - 1200°C (932 - 2192°F)
Spectral range 1,4 - 1,75μm and 1,6 - 1,75 μm
Response time (t95) <20 ms
Measuring uncertainity 0,5 % of meas. value in °C ± 1°C (ε=1, Tu=23°C, T95 = 1s)
Reproducibility 0,1 % of meas. value in °C ± 1°C (ε=1, Tu=23°C, T95 = 1s)
Emission ratio 0,8 - 1,2 adjustable at the unit or through interface
Emission factor ε 100 - 10 % through interface
Analog output 0 - 20 mA oder 4 - 20 mA, load max. 500R
Part measuring ranges: free adjustable within the measuring range
Resolution <0,1% analog output, <0,1°C at interface
1 limit output (open coll.) 24 V DC / max. 100 mA
Max. reading memory max.memory, erasing after time, external contact, by software, double memory
Interface RS 232 ± 50 V isolated
Software IR-LOG data recording, graph. representation, setting of pyrometer parameters
adjustable parameter
with Software IR-LOG
emissionfactor,switching output,analog output, part meas.range, °C / °F,
max.memory, average value, light beam aiming device switchable
Objectives for accommodation to the measuring application an extensive selection
of objectives are available
Working temperature pyrometer 0 - 50°C (32 - 122°F), optic system up to 150°C (302°F)
Stock temperature -10°C - +70°C (14 - 158°F)
Temperature sensitivity 0,05 % / °C
Humidity tolerance 35 - 85 % RF (non condensing)
Operating Voltage 24 V DC ± 10 % or 18 V AC ± 10 % < 160 mA
Unit connection 12-pole plug-connector
Dimensions H / W / D 54 x 54 x 147 mm (2,13 x 2,13 x 5,79 inch) / ALU-case
Weight 0,6 kg (1,32 lbs)
Protection grade IP 65


Datasheet QKTRD 1075
Datasheet


mechanical assembly electrical assembly electronic process unit
execution in cooling case connection cable 12-pole AE 1010 electr. process unit
Blowing device power supply 230 VAC - 24 VDC AE 1012 electr. process unit
Mirror 90° digital display (built in-execution)
Mounting parts line scanner SC 1000 / 1012


back